Chemistry:New mode for the on-line determination of oxonium ion in differentstrong acids using CFIA via the use of homemade linear array

New mode for the on-line determination of oxonium ion in differentstrong acids using CFIA via the use of homemade linear array Ayah 5Sx4-ST- 5D solar CFI analyser

Issam M. A. Shakir*, and Hussein F. A. AL-Razack

Department of chemistry, College of science, University of Baghdad, Baghdad, Iraq.

Abstract

Strong acids were determined via the blue complex solution reaction of a loaded ferric (III) ion on strong cation exchange resin which in turn reacts with potassium hyxacyano ferrate (II). The attenuation effect of formed blue complexFe4([Fe(CN)6])3on (0 -180o) incident LED light was measurement via homemade linear array AYAH 5SX4-ST-5D solar CFI analyser. Optimum parameters were 0.005M.L-1  [Fe(CN)6]-4  , flow rate of 3mL.min-1 , sample volume 204 µL , sample purge time of 70 seconds was chosen, and 1V for light intensity. A liner calibration graph of 0.005 -0.15 M.L-1 were obtains for HCl, HNO3, HCLO4 and H2SO4, with a linearity (r2 %) 94 -98 % and L.O.D based on gradual dilution of lowest concentration in calibration graph was 22.314 µg for HCl, 38.563 µg for HNO3,61.480 µg for HCLO4, and 60.020 µg for H2SO4 .Repeatability (RSD) 0.5% was obtained. the proposed method was compared with standard method. turbidity measured using HANNA instrument for available acid in commercial (HCl, H2SO4). 

طريقة جديدة انيه لتقدير ايون الاوكسونيوم في الحوامض القويه المختلفة باستخدام تقنيه التحليل بالحقن الجرياني المستمر من خلال استخدام محلل الحقن الجرياني المستمرlinear array Ayah 5Sx4-ST- 5D solar

عصام محمد علي شاكر * , حسين فارس عبد الرزاق

قسم الكيمياء, كلية العلوم, جامعة بغداد , بغداد , العراق.

الخلاصة:

قدرت الحوامض القوية عن طريق محلول معقد ازرق والمعتمد على التفاعل بين ايون الحديد الثلاثي والمحمل على مبادل ايوني حامضي قوي وبوناسيوم سداسي ثايوسيانات الحديد (II). قياس توهين الضوء المسلط بزاوية ( 0-180˚) والمعتمد على تكوين معقد ازرق Fe4([Fe(CN)6])3 يتم عن طريق محلل الحقن الجرياني المستمر والمصنع محليا  linear array AYAH 5SX4-ST-5D solar CFI analyser. تم اختيار الظروف الفضلى  0.005مول.لتر-1[Fe(CN)6]-4  وسرعة الجريان 3 مل. دقيقة-1  ,حجم الأنموذج المحقن 204مايكروليتر ,زمن حقن الأنموذج 70 ثانية ويالأضافة الى 1 فولت وهي شدة الضوء الساقط . منحني معايرة بحدود 0.005 – 0.15 مول .لتر-1 لكل من HCl, HNO3, HClO4 بالأضاقة الى  H2SO4مع نسبة خطيه (r2% ) تمتد من 94- 98 % بالأضافة الى  حدود الكشف (LOD ) والمستندة على التخفيف التدريجي لأقل تركيز في منحني المعايرة 22.314مكغم لأنموذج HCl , 56.38مكغم لـ HNO3, 48.61 مكغم لـ HClO4بالأضافة الى 02.60مكغم  لـ H2SO4 . تم الحصول على تكرارية (RSD% ) 0.5% .تم مقارنة الطريقة المقترحة مع الطريقة الكلاسيكية(قياس التعكريه) باستخدام جهازHANNA  لكل من الحوامض المتوفرة (HCl , H2SO4).

 

 

 

 

alt

New mode for the on-line determination of oxonium ion in differentstrong acids using CFIA via the use of homemade linear array Ayah 5Sx4-ST- 5D solar CFI analyser

 

Issam M. A. Shakir*, and Hussein F. A. AL-Razack

Department of chemistry, College of science, University of Baghdad, Baghdad, Iraq.

 

Abstract

Strong acids were determined via the blue complex solution reaction of a loaded ferric (III) ion on strong cation exchange resin which in turn reacts with potassium hyxacyano ferrate (II). The attenuation effect of formed blue complexFe4([Fe(CN)6])3on (0 -180o) incident LED light was measurement via homemade linear array AYAH 5SX4-ST-5D solar CFI analyser. Optimum parameters were 0.005M.L-1  [Fe(CN)6]-4  , flow rate of 3mL.min-1 , sample volume 204 µL , sample purge time of 70 seconds was chosen, and 1V for light intensity. A liner calibration graph of 0.005 -0.15 M.L-1 were obtains for HCl, HNO3, HCLO4 and H2SO4, with a linearity (r2 %) 94 -98 % and L.O.D based on gradual dilution of lowest concentration in calibration graph was 22.314 µg for HCl, 38.563 µg for HNO3,61.480 µg for HCLO4, and 60.020 µg for H2SO4 .Repeatability (RSD) 0.5% was obtained. the proposed method was compared with standard method. turbidity measured using HANNA instrument for available acid in commercial (HCl, H2SO4). 

 

Keywords: Turbidity, Oxonium ion, ion exchange

 

طريقة جديدة انيه لتقدير ايون الاوكسونيوم في الحوامض القويه المختلفة باستخدام تقنيه التحليل

بالحقن الجرياني المستمر من خلال استخدام محلل الحقن الجرياني المستمر

linear array Ayah 5Sx4-ST- 5D solar

 

عصام محمد علي شاكر * , حسين فارس عبد الرزاق

قسم الكيمياء, كلية العلوم, جامعة بغداد , بغداد , العراق.

 

الخلاصة:

قدرت الحوامض القوية عن طريق محلول معقد ازرق والمعتمد على التفاعل بين ايون الحديد الثلاثي والمحمل على مبادل ايوني حامضي قوي وبوناسيوم سداسي ثايوسيانات الحديد (II). قياس توهين الضوء المسلط بزاوية ( 0-180˚) والمعتمد على تكوين معقد ازرق Fe4([Fe(CN)6])3 يتم عن طريق محلل الحقن الجرياني المستمر والمصنع محليا  linear array AYAH 5SX4-ST-5D solar CFI analyser. تم اختيار الظروف الفضلى  0.005مول.لتر-1[Fe(CN)6]-4  وسرعة الجريان 3 مل. دقيقة-1  ,حجم الأنموذج المحقن 204مايكروليتر ,زمن حقن الأنموذج 70 ثانية ويالأضافة الى 1 فولت وهي شدة الضوء الساقط . منحني معايرة بحدود 0.005 – 0.15 مول .لتر-1 لكل من HCl, HNO3, HClO4 بالأضاقة الى  H2SO4مع نسبة خطيه (r2% ) تمتد من 94- 98 % بالأضافة الى  حدود الكشف (LOD ) والمستندة على التخفيف التدريجي لأقل تركيز في منحني المعايرة 22.314مكغم لأنموذج HCl , 56.38مكغم لـ HNO3, 48.61 مكغم لـ HClO4بالأضافة الى 02.60مكغم  لـ H2SO4 . تم الحصول على تكرارية (RSD% ) 0.5% .تم مقارنة الطريقة المقترحة مع الطريقة الكلاسيكية(قياس التعكريه) باستخدام جهازHANNA  لكل من الحوامض المتوفرة (HCl , H2SO4). Normal 0 false false false EN-US X-NONE AR-SA
 